If you need to separately count alpha and beta concurrently and log the results, the Model 2360 will be the right tool. This instrument is an optimized Alpha-Beta Ratemeter, Scaler, and Data Logger that offers a traditional analog readout supplemented with a digital LCD for the scaler function. This meter supports many different externally connected detectors; two of the more popular ones are noted below.
Up to 550 data points can be stored. Each data point consists of a sample number, date/time stamp, both alpha and beta sample measurements, and a 10 character location identifier. The location identifier can be input via an optional external device through the RS-232 serial port.
A front panel mounted three-way switch facilitates selecting alpha only, beta only, or alpha plus beta. Other controls include a rotary switch for selecting any of the four-decade ranges, battery test, and instrument shut-off; a switch for viewing the high voltage or resetting the counts; an audio volume adjustment knob; a scaler count time selector rotary switch; and a handle mounted push-button for logging and starting scaler count cycles.
Capable of storing up to 550 individual data points with the following identifiers for each point (all data is stored in non-volatile memory allowing batteries to be removed without loss of data):
Barcode Scanner for Model 2360
PC interface software and the driver needed to use the RS-232 to USB adapter cable are available for download on the Software & Firmware Downloads page.